Associated Brands

nCompass™ Series Patient Monitors

nCompass™ Series Patient Monitors

Features

  • Easy-to-read 12.1” touch screen display
  • Easy to learn, easy to use simplified menus
    and dedicated function keys.
  • Mini-Trends can display condensed tabular
    vital signs data to show recent monitoring
    activities.
  • OXY-CRG Trend Display.
  • Optional invasive cardiac output interface
    capability
  • Compatible with Criticare’s 8500 Series
    Ag e nt Gas Analyzer for anesthetic
    monitoring.

Improve patient outcomes with a comprehensive, user-configurable
patient monitoring system covering the needs of multiple
departments within hospitals and outpatient facilities.

Configurations

  • STANDARD: ECG, ComfortCuff™
    DOX™ SpO2 , HR, Temperature, Respiration,
    Arrhythmia and ST analysis and Nurse Call
  • OPTIONAL: CO 2, 1 or 3 channel IBP, printer.
  • Focus on your patients, not your monitor.

    The nCompass™ patient monitor is the perfect choice for busy departments demanding accurate, reliable performance.

    The nCompass™ can monitor adult and pediatric patients in virtually every department. The innovative user interface features simplified menus and dedicated function keys for timely setups. The big, bright display with large numer ics can be easily seen from across the room.

    nCompass™ monitors mount quickly and easily to walls and rollstands.

    ECG
    Number of Leads: 3- or 5-Lead
    Lead View: User-selectable; I, II, III, aVR, aVL, aVF, V (5-Lead);
    I, II, or III (3-Lead)
    Gain Selection: x0.5, x1, x2, and x4
    Frequency Response: Diagnostic: 0.05 to 100 Hz (-3dB);
    Monitor: 0.5 to 40 Hz (-3dB);
    ST: 0.05 to 40 Hz (-3dB)
    Electrosurgery Protection: Yes
    Defibrillator Protection: Yes
    Pacer Detection/Rejection: Yes
    ST Accuracy: ±0.2 mm or ±5% of reading, whichever is greater,
    over heart range of 30 to 120 bpm
    ST Range: -6.0 to 6.0 mm
    NIBP (Non-Invasive Blood Pressure)
    Technique: Oscillometric upon inflation
    Measurement Cycle: < 40 sec. typical
    Auto. Measurement Cycles 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
    Range: Adult/Pediatric: 30 to 300 mmHg;
    Neonate: 20 to 150 mmHg
    Resolution: 1 mmHg
    Transducer Accuracy: ±2 mmHg or ±2% of reading, whichever is greater,
    over full range
    STAT Mode: 5 min of consecutive readings
    Clinical Accuracy: Complies with AAMI SP-10:2002
    Pulse Oximetry
    Range: 1% to 99%
    Resolution: 1%
    Accuracy: 70% to 99% range: ±2%; 50% to 69% range: ±3%;
    < 50%: unspecified
    Method: Dual wavelength LED
    Heart (Pulse) Rate
    Source: User selectable: Smart, ECG, Pleth, NIBP, IBP
    Range: 20 to 300 bpm (ECG, Pleth);
    30 to 240 bpm (NIBP); 30-300 bpm (IBP)
    Accuracy: ±1 bpm or ±1% (ECG) whichever is greater
    Temperature
    Channels: 2
    Probe: YSI-700 Series
    Range, Accuracy: 68° to 113°F (20° to 45°C): ±0.1°C
    Probe: YSI-400 Series
    Range, Accuracy: 59° to 113°F (15° to 45°C): ±0.1°C
    Display Resolution: ±0.1°
    Respiration
    Rate: 6 to 150 bpm (ECG); 2 to 120 bpm (CO2 )
    Resolution: 1 bpm
    Accuracy: ±2 bpm or ±2%, whichever is greater
    Source: User selectable: Smart, CO2 , ECG
    CO2
    Units: mHg; %; kPa; Torr
    Method: Non-dispersive infrared; autocalibrating
    N 2O Compensation: ±2 bpm or ±2%, whichever is greater
    Range: 0 to 99 mmHg; 0 to 12.5%; 0 to 12.5 kPa; 0 to 99 Torr
    Resolution: 1 mmHg; 0.1%; 0.1 kPa; 1 Torr
    Flow, Sample Rate: 200 ml/min
    Accuracy: ±2 mmHg; ±0.3 vol. %; ±0.3 kPa; ±2 Torr
    or ±4% of reading, whichever is greater,
    at 200 ml/min & respiratory rate of ≤ 120 Br/min
    Rise Time: 170 ms (10 to 90%) at 200 ml/min
    IBP (Invasive Blood Pressure)
    Channels: 1 or 3
    Transducer Sites: Selectable: ART, PA, LV, RV, LA, RA, CVP, ICP
    Pressure Range: -10 mmHg to 300 mmHg
    Transducer Freq. Response: 0 to 20 Hz; Impedance >300 ohms;
    Transducer Sensitivity: 5μV/V/mmHg
    Accuracy: ±1 mmHg or ±1%, whichever is greater
    Alarms
    Indication: Audible; Visual (EN 60601-1-8)
    Silence: Yes; 2 minutes or permanent
    Trends
    Memory: 24 hours at 30 seconds intervals
    Display: Tabular, Graphical
    Display
    Screen: 12.1” acive color touch screen
    Resolution: Internal: 800 X 600 pixel
    External: 800 X 600 pixel
    Waveform: 6, maximum
    Waveform Types: ECG Leads I, II, III, aVR, aVL, aVF, V; CO2 ;
    Respiration; Pleth; IBP; breath-bybreath CO2
    ST histogram; Mini-Trends; OXY-CRG;
    Sweep Speed: 6.25, 12.5, 25, or 50 mm/sec
    Controls
    Keys: 10; membrane-activated
    Rotary Knob: Push and rotate; 24 steps/turn
    System Outputs
    COM ports: Two RS 232 compatible Db9 ports;
    USB (for software downloads only)
    Nurse Call 5.67 in (H) x 7 in (W) x 4.81 in (D) 14.4 cm (H) x 17.8 cm (W) x 12.2 cm (D)
    Defib/Sync: BNC connector
    Video Port: Serial VGA compatible
    DC Out: 5 VDC output for optional wireless accessory
    External Module Port: 8-PIN circular connector
    Printer (optional)
    Type: Internal thermal line
    Paper Speed: 12.5, 25, or 50 mm/sec
    Physical Properties
    Weight: 12 lbs; 5.4 kg
    Size: 12” (H) x 14.5” (W) x 11.5” (D);
    30.5 cm (H) x 36.8 cm (W) x 29.2 cm (D)
    Power Requirements
    Voltage: 100 to 240 VAC; 50/60 Hz
    Power Consumption: 35 W, typical
    Battery: 1 sealed lead acid
    Battery Life: 2 hours, typical (3 hours, typical for base unit)
    Recharge Time: 4 to 6 hours, typical
    Environmental
    Operating Temperature: 32° to 104°F (0° to 40°C)
    Storage Temperature: 23° to 122°F (-5° to 50°C
    Operating Humidity: 15% to 90%, noncondensing
    Operating Altitude: -300 to 3,000 meters
    Classification
    Medical Device: Class II Equipment (IIb EU)
    Electrical Protection: Class I Equipment
    Degree of Protection: Type CF, Defibrillator-proof
    Protection Against Ingress: IPX1
    Safety: EN 60601-1

    For further information , you can  Download Brochure   or    Contact Sales