Associated Brands

Poet® IQ Anesthetic Gas Analyzer

eQuality™ Patient Monitor

Features

  • State-of-the-art non-dispersive infrared (NDIR) technology identifies and measures five anesthetic agent gases: Halothane, Enflurane, Isoflurane Desflurane and Sevoflurane.
  • Automatic or manual modes.
  • Breath-by-breath O2, CO2, and N2O monitoring.
  • Gas flow sampled at 100, 150 and 200 ml/min.
  • Optional printer, with user - definable report formats.
  • Mixed agent identification.
  • Fast warm-up time ensures full accuracy
    within minutes.
  • Auto-calibration.
  • Bright, real-time display of numerical values and waveforms provides instant notification of changing patient status.
  • Lightweight, portable design provides
    flexible workspace options.

Description

Criticare’s Poet ® IQ Anesthetic Gas Analyzer, when paired with the nCompass™ Patient Monitor, provides a unique combination of leading edge vital signs technology and anesthesia gas monitoring in a compact, modular system.

A powerful combination—proprietary CSI vital signs and 5-agent gas analysis technology. Versatile, reliable, and incredibly compact. Designed for anesthesia applications in hospitals and outpatient surgical centers.

The Poet®IQ automatically identifies and quantifies inspired and expired O2, CO2, N2O, and five anesthetic agents. The system’s reliable performance, ease of use, flexible design, and affordable cost make it the ideal monitoring solution for anesthesia applications in hospitals and surgical centers.

Technical Specifications

Gas Monitoring
Method Sidestream; Non-dispersive infrared (NDIR)
Identified Gases Halothane, Enflurane, Isoflurane, Desflurane,
Sevoflurane, O2, CO2, N2O
Concentration Units Vol%, Torr, kPa, mmHg
Flow Rates 100, 150 or 200 ml/min
Agent Detection
Measurement Range Halothane: 0 to 10%
Enflurane: 0 to 10%
Isoflurane: 0 to 10%
Desflurane: 0 to 20%
Sevoflurane: 0 to 10%
CO2: 0 to 12.5%
N2O: 0 to 99%
O2: 0 to 100%
Measurement Accuracy Agents: ±0.1% abs. + 4% of reading
CO2: ±0.2% abs. or 4% of reading
N2O: ±1.5% abs. + 4% of reading
O2: ±3 vol% (0 – 90%)
±4 vol% (91 – 99%)
Time to Detect Agent < 15 seconds @ 200 ml/min
Agent Detection
Resolution
0.1 vol%
Mixed Gas Threshold 0.2 vol% +10% of total concentration
Rise Time Agents: 450 msec
CO2: 350 msec
N2O: 400 msec
O2: 600 msec
Respiration Rate
Range 1 to 60 Br/min
Accuracy ±2 Br/min or 2% of reading
System Features
Occlusion Clearing Automatic
Auto Zeroing Occurs 30 to 60 minutes
Duration: 3.0 to 7.0 seconds
Manual user calibration not required.
Temperature stabilized optical assembly.
Auto-calibration; verification
recommended once per year.
Warm-up Time 1 minute to first waveforms;
< 20 minutes to full accuracy
Alarms
Alarm Characteristics EN 475, Adjustable; with audible and visual
indications from the patient monitor.
Alarm Levels High, Medium, Low, Informational
Alarm Modes Adult/Pediatric/Neonate
High and low limit settings for each mode.


Controls
Memory 24 hours of stored data in patient monitor
Display Tabular, Graphical
Display Connects to nCompass™
8100H Series patient monitors
Languages Consult with sales or customer service for
available language configurations
System Outputs
System Configuration Modular design with bi-directional communication,
via cable, to the host monitor.
Waveform Output Halothane, Enflurane, Isoflurane, Desflurane,
Sevoflurane, CO2, N2O, O2
Output Data Inspired and end-tidal gas concentrations;
continuous real-time gas concentrations;
respiratory rate (elapsed time since last breath);
agent identification (primary agent and
mixed agents); system information (diagnostic
status messages)
Power Requirements
Voltage: 12 VDC, typical
Receives power from host monitor
Power Consumption: 6W peak, 3W typical
Mechanical
Weight: 5.9 lbs. (2.7 kgs.)
Size: Height: 4.0 in (10.2 cm)
Width: 14.0 in (35.6 cm)
Depth: 10.8 in (27.3 cm)
Environmental
Operating Temperature: 59° to 95°F (15° to 35 °C)
Storage Temperature: 23° to 122°F (-5° to 50 °C)
Operating/Storage
Humidity:
15% to 95%, noncondensing
Altitude: -1,000 ft to 10,000 ft (-300 m to 3,000 m)
Classification
Medical Device Class II Equipment (IIb EU)
Electrical Protection Class I Equipment
Degree of Protection Type CF, Defibrillator-Proof
Protection Against
Ingress
Ordinary

For further information , you can  Download Brochure   or    Contact Sales